Assessment of Radiofrequency Device Sensitivity for the Detection of Retained Surgical Sponges in Patients With Morbid Obesity

Victoria M. Steelman, PhD; Mohammad H. Alasagheirin, MSN, RN
Arch Surg. 2012;147(10):955-960. doi:10.1001/archsurg.2012.1556.
Text Size: A A A
Published online

Hypothesis  Retained surgical sponges are serious medical errors that result in negative patient outcomes. A radiofrequency (RF) mat for the operating room bed has recently been introduced to detect the presence of a retained surgical sponge. The study objectives were to evaluate the sensitivity and specificity of the RF mat for the detection of surgical sponges through the torso of individuals with varying body habitus and to compare the sensitivity of the RF mat with that of the RF wand. We hypothesized that the sensitivity and specificity of the RF technology would be comparable to published findings of the manual sponge count by operating room personnel.

Design  A prospective, crossover, double-blinded study design was used. Participants served as their own controls.

Setting  Large Midwestern academic medical center.

Participants  In total, the first phase of the study enrolled 203 participants, including 129 (63.5%) with morbid obesity. One hundred seventeen of 203 participants were also enrolled in the second phase of the study.

Main Outcome Measures  The study participants reclined in a supine position on top of an RF mat. Four surgical sponges were sequentially placed on top of the torso in locations approximating the abdominal quadrants. The torso was scanned for sponges. In a subset of participants, 4 surgical sponges were sequentially placed underneath the torso, and an RF wand was passed over the abdomen.

Results  Overall, 812 readings were obtained with the RF mat, and 468 readings were obtained with the RF wand. Twelve false-negative readings were obtained with the RF mat, exclusively in participants with super morbid obesity (body mass index [calculated as weight in kilograms divided by height in meters squared] >50.0). Overall, the sensitivity of the RF mat was 98.1%, and the specificity of the RF mat was 100.0%. In the subset of 117 participants in whom the RF wand was also used, the sensitivity and specificity of the wand were each 100.0%.

Conclusions  The sensitivity and specificity of RF device technology are much higher than those of surgical sponge counts or published findings on the use of intraoperative radiographs to identify retained surgical sponges. The RF wand is more sensitive than the RF mat in individuals with morbid obesity.

Sign In to Access Full Content

Don't have Access?

Register and get free email Table of Contents alerts, saved searches, PowerPoint downloads, CME quizzes, and more

Subscribe for full-text access to content from 1998 forward and a host of useful features

Activate your current subscription (AMA members and current subscribers)

Purchase Online Access to this article for 24 hours

Figures

Tables

Interactive Graphics

Video

Country-Specific Mortality and Growth Failure in Infancy and Yound Children and Association With Material Stature

Use interactive graphics and maps to view and sort country-specific infant and early dhildhood mortality and growth failure data and their association with maternal

References

Correspondence

CME


You need to register in order to view this quiz.
NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s “Cited By” API will populate this tab (http://www.crossref.org/citedby.html).
Submit a Comment

Some tools below are only available to our subscribers or users with an online account.

Sign In to Access Full Content

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Topics
PubMed Articles
Jobs